Accuracy and Reproducibility of a Modified Echocardiographic Method for Right Ventricular Output Calculation in Neonates

Carlotta Milocchi, Silvia Nogara, Giorgia Mazzuca, Federica Runfola, Martina Ciarcià, Iuri Corsini, Benjamim Ficial

We aimed to evaluate the accuracy and reproducibility of right ventricular output (RVO) using different anatomical landmarks: the internal pulmonary valve diameter (PVD) between the valve hinge points (hinge-PVD) according to the traditional technique, and PVD between the valve leaflet tips (tip-PVD). This was a retrospective analysis of prospective collected data. All neonates with echocardiographic measurements of RVO and left ventricular output (LVO) without congenital heart disease, including patent ductus arteriosus and patent foramen ovale > 3 mm, were included. Accuracy was assessed by comparison with LVO. Intra- and inter-observer reproducibility of the off-line analysis were assessed. Forty-five neonates were included. RVO calculation with tip-PVD was more accurate than hinge-PVD in comparison with LVO, r2 0.712 versus 0.464, bias (95% limits of agreement) 1.4 mL/kg/min (−26–29 mL/kg/min) versus 61 mL/kg/min (−11–132 mL/kg/min), respectively. Both hinge-PVD and tip-PVD presented similar reproducibility, with an intra-observer bias (95% LOA) of 0.3 (−1.0–0.5) and –0.2 (−0.8–0.5) respectively, and an inter-observer bias of 0.1 (−1.3–1.6) and 0.1 (−1.4–1.6). RVO calculation using tip-PVD was more accurate than the conventional technique, with similar reproducibility.