DOI: 10.1177/01617346241311901 ISSN: 0161-7346

Clinical Feasibility of 3-D Acoustic Radiation Force Impulse (ARFI) Imaging for Targeted Prostate Biopsy Guidance

Derek Y. Chan, Spencer R. Moavenzadeh, Wren E. Wightman, Mark L. Palmeri, Thomas J. Polascik, Kathryn R. Nightingale

We have developed a 3-D acoustic radiation force impulse (ARFI) prostate imaging system to identify regions suspicious for cancer and guide a targeted prostate biopsy in a single visit. The system uses a side-fire transrectal probe and an automated rotation stage to acquire ARFI and B-mode image volumes, combined with 3-D visualization and targeting software to enable biopsy target identification and guide a transperineal (TP) biopsy. The system was tested in the first clinical trial of its kind, with subjects serially undergoing ARFI-guided targeted TP biopsy, multiparametric magnetic resonance imaging (mpMRI)-ultrasound fusion TP biopsy, and systematic sampling TP biopsy. The findings indicate that the ARFI system is feasible for guiding a targeted biopsy. For lower-grade cancer (grade groups [GG] 1–2), ARFI underperformed mpMRI and systematic sampling, detecting cancer in 54% of GG 2 subjects. However, ARFI performance improved with increasing cancer grade; for higher-grade cancer (GG 3–5), ARFI performed comparably to the other biopsy approaches, and detected cancer in all GG 4 and 5 subjects. The findings also suggest the benefit of using 2-D ARFI imaging to confirm target location during live B-mode imaging, which could improve existing ultrasonic fusion biopsy workflows. This study is registered with ClinicalTrials.gov as NCT04607135.

More from our Archive